| Probe Station Setup |
|
|
|
configuration
|
|
up to 4 independent probes, fixed sample |
| advantages |
|
compact, simple setup, fixed sample HF connection
possible |
| |
|
|
|
| Sample and Probe Positioning |
|
|
|
sample positioning |
|
fixed sample |
sample size |
|
< 13 mm diameter |
probe positioning |
|
each probe mounted on an independent
ANPxyz51 positioning stage |
number of probes |
|
up to 4 probes, each accessible
in xyz
|
probe travel range |
|
4 x 4 x 3 mm³ manipulation
space for each probe |
minimum step size |
|
approx. 50 nm @ 300 K, 10 nm @ 4
K |
| |
|
|
|
| Probes |
|
|
|
DC probes |
|
double tip probes, GS type, ideal
for DC biasing |
|
|
standard pitch
50 µm (others
on request) |
HF probes standard |
|
GSG probes up to 5 GHz (others on
request), |
|
|
standard pitch
50 µm (others
on request) |
fiber probes |
|
single/multi mode fibers for localized
illumination |
| |
|
|
|
| Sample Stick |
|
|
|
DC wiring / feed-throughs |
|
up to 40 contacts
for positioners, temperature sensor, LED (additional DC
wiring on request) |
HF wiring |
|
4 coaxial HF lines
including SMA feed-throughs (more on request) |
length |
|
1 .. 1.5 m, customized |
diameter
|
|
2 " (50.8
mm) |
sample monitoring |
|
sample / probe monitoring via free space
optics and |
|
|
CCD camera, resolution
approx. 5 - 10 µm, |
|
|
field of view
max. 5 - 6 mm, magnification to CCD chip max. 1:10 |
sample illumination |
|
infrared LED (others
on request) |
| |
|
|
|
| Operating Conditions |
|
|
|
temperature range |
|
1 .. 300 K (dependent on cryostat) |
magnetic field range
|
|
0 .. 15 T + (dependent on magnet) |
operating pressure range |
|
1E-6 mbar .. 1 bar (designed for
exchange gas atmosphere) |
| |
|
|
|
| Cooling Specifications |
|
|
|
| bore size |
|
designed for a 2" (50.8 mm)
cryostat/magnet bore |
| cryostat |
|
LTSYS-He4, LTSYS-Cc |
| |
|
|